6 set up, 1 general, 2 adjustment elements – VEGA VEGAVIB 65 - relay output User Manual

Page 18

Advertising
background image

6 Set up

6.1 General

The numbers in brackets refer to the following illustrations.

On the electronics module you will find the following indicating

and adjustment elements:

l

Potentiometer for switching point adaptation (1)

l

DIL switch for mode adjustment - min./max. (2)

l

Signal lamp (5)

Note:

As a rule, always set the mode with mode switch (2) before

starting the setup of VEGAVIB 65 . The switching output will

change if you set the mode switch (2) afterwards. This could

possibly trigger other connected instruments or devices.

6.2 Adjustment elements

3

4

5

1

2

Fig. 11: Oscillator VB60R - Relay output

1

Potentiometer for switching point adaptation

2

DIL switch for mode adjustment

3

Ground terminal

4

Screwed terminals

5

Control lamp

Function/Configuration

18

VEGAVIB 65 - with relay output

Set up

29283

-EN
-060508

Advertising