VEGA VEGASWING 61 - Relay (DPDT) User Manual

Page 28

Advertising
background image

28

9 Supplement

VEGASWING 61 • - Relay (DPDT)

29224-EN-130409

Ambient conditions
Ambient temperature on the housing

-40 … +70 °C (-40 … +158 °F)

Storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions
Measured variable

Limit level of liquids

Process pressure

-1 … 64 bar/-100 … 6400 kPa (-14.5 … 928 psig)
The process pressure is dependent on the process

fitting, for example Clam p or flange (see the following

diagrams)

Process pressure - Clamp fittings

Ʋ Clamp connection up to 2"

-1 … 16 bar/-100 … 1600 kPa (-14.5 … 232 psig)
max. 150 °C (302 °F)

Ʋ Clamp connection > 2"

-1 … 0.5 bar/-100 … 50 kPa (-14.5 … 7.25 psig)
> 150 °C (> 302 °F)

Maximum allowable operating pressure 100 bar/10000 kPa (1450 psig) or 1.5 times process

pressure
The function of the instrument is ensured up to an

operating pressure of 100 bar/10000 kPa (1450 psig)

at a maximum process temperature of +50 °C (+122 °F)

(only with threaded versions).

Process temperature (thread or flange

temperature)

Ʋ VEGASWING 61 of 316L/Hastelloy

C22 (2.4602)

-50 … +150 °C (-58 … +302 °F)

Process temperature (thread or flange

temperature) with temperature adapter

(option)

Ʋ VEGASWING 61 of 316L/Hastelloy

C22 (2.4602)

-50 … +250 °C (-58 … +482 °F)

Ʋ VEGASWING 61 enamelled

-50 … +200 °C (-58 … +392 °F)

Ʋ VEGASWING 61 with ECTFE coating -50 … +150 °C (-58 … +302 °F)

Ʋ VEGASWING 61 with PFA coating

-50 … +150 °C (-58 … +302 °F)

Advertising