3
.2 Principle of operation
VEGAFLEX
65
is a level sensor with pipe probe for continuous level
measurement.
I
t is designed for industrial use in all areas of process technology and
can be used in liquids.
H
igh frequency microwave pulses are guided along a steel rope or a
rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and
outputted as distance.
F
our-wire electronics with separate power supply.
T
he supply voltage range can differ depending on the instrument
version.
T
he data for power supply are specified in chapter "Technical data".
M
easured value transmission is carried out via the 4 … 20 mA/HART
output separate from power supply.
T
he background lighting of the indicating and adjustment module is
powered by the sensor. A certain level of operating voltage is required
for this. You can find the exact voltage specifications in chapter
"T
echnical data".
3
.3 Operation
T
he instrument can be adjusted with the following adjustment media:
l
W
ith indicating and adjustment module
l
with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC
l
with manufacturer-specific adjustment programs AMS™ or PDM
l
W
ith a HART handheld