VEGA VEGACAL 67 Foundation Fieldbus User Manual

Page 38

Advertising
background image

38

10 Supplement

VEGACAL 67 • Foundation Fieldbus

31760-EN-130917

Ʋ with external housing

-50 … +400 °C (-58 … +752 °F)

2

1

3

200 °C

(292 °F)

300 °C

(572 °F)

100 °C

(212 °F)

400 °C

(752 °F)

0 °C

(32 °F)

-50 °C

(-58 °F)

16 bar

(232psi)

Fig. 20: Process temperature - Process pressure
1 Process temperature

2 Process pressure

3 Temperature range with external housing

Dielectric constant

≥ 1.5

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

3)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap M20 x 1.5; 1 x blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; 1 x blind plug M16 x 1.5 or

optionally available with 1 x plug M12 x 1 for external

display and adjustment unit

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, 1 x blind

plug M16 x 1.5 or optionally 1 x plug M12 x 1 for exter-

nal display and adjustment unit

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; 1 x blind plug M16 x 1.5 or optionally avail-

able with 1 x plug M12 x 1 for external display and

adjustment unit

Spring-loaded terminals for wire cross-

section

< 2.5 mm² (AWG 14)

Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry

Ʋ Single chamber housing

1 x IP 68 cable gland M20 x 1.5; 1 x blind plug M20 x 1.5

Ʋ Double chamber housing

1␣x IP␣68 cable gland M20␣x␣1.5; 1␣x blind stopper

M20␣x␣1.5; 1␣x blind stopper M16␣x␣1.5

3)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/8" FF.