VEGA VEGAFLEX 67 (-20…+250°C) Profibus PA User Manual
Page 49
L
0
10 mm (0.394
")
-10 mm (-0.394
")
0,1 m (3.937
"
)
0,15 m
(5.906
")
1
1
Fig. 24: Deviation VEGAFLEX 67 in coax version
1
Dead zone - no measurement possible in this area
L
Probe length
Influence of the ambient temperature to the sensor electronics
Temperature drift
0.03 %/10 K relating to the max. measuring range
or max. 0.3 %
Temperature drift - Digital output
3 mm/10 K relating to the max. measuring range or
max. 10 mm
Ambient conditions
Ambient, storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Process conditions
Process pressure
-1 … +100 bar/-100 … +10000 kPa (-
-14.5 … +1450 psig), depending on the process
fitting
Process temperature (thread or flange tem-
perature)
-20 … +250 °C (-4 … +482 °F)
The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.
VEGAFLEX
67 • Profibus PA
49
10 Supplement
32309
-
EN
-120329