4 recurring function test – VEGA VEGAVIB 66 - NAMUR User Manual

Page 22

Advertising
background image

Note:

The mode adjustment on the NAMUR amplifier must be

selected in such a way that the switching output goes to safe

position in case of a failure message (I <=1 mA).

Level

Signal current -

Sensor

Control lamp

Falling characteris-

tics max.

>=2.2 mA

Falling characteris-

tics max.

<= 1 mA

Rising characteris-

tics min.

>=2.2 mA

Rising characteris-

tics min.

<= 1 mA

Failure

any

<= 1 mA

flashes

6.4 Recurring function test

According to IEC 61508.
VEGAVIB 66 is qualified in mode A (overfill protection) for use

in measuring chains of stage SIL2 according to IEC 61508

(redundant, stage SIL3).

The following instrument combinations meet the requirements

according to SIL:
VEGAVIB 66

l

Oscillator VBE60N

The recurring test according to IEC 61508 can be carried out

by pushing the simulation key on the oscillator or by briefly (>2

seconds) interrupting the cable to the sensor. The correct

SIL

Recurring function test

22

VEGAVIB 66 - with NAMUR output

Set up

31219

-EN
-060508

Advertising