VEGA VEGAFLEX 62 4 … 20 mA_HART four-wire User Manual
Page 13
≥ 300 mm (11
13
/
16
")
150 mm (5
29
/
32
")
150mm (5
29
/
32
")
1
SW 13
L
Fig. 3: Bow-shaped segment ø 16 mm (0.63 in)
1
Max. filling height
L
150 mm (5.91 in)
B
efore beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
W
ith threaded versions, the housing must not be used to screw in the
instrument! Applying tightening forces on the housing can damage its
internal parts.
U
se the hexagon for screwing in.
U
se the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
Y
ou can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This
applies mainly to outdoor mounting as well as installation in areas
where high humidity is expected (e.g. through cleaning processes) or
on cooled or heated vessels.
Welding work
Handling
Moisture
VEGAFLEX
62 • 4 … 20
mA/HART four-wire
13
4 M
ounting
31838
-
EN
-100426