7 Maintenance and fault rectification

VEGACAL 67 • For connection to a signal conditioning instrument

31761-EN-130917

7 Maintenance and fault rectification

7.1 Maintenance

If the instrument is used properly, no special maintenance is required

in normal operation.

7.2 Rectify faults

The operator of the system is responsible for taking suitable meas-

ures to rectify faults.

VEGACAL 67 offers maximum reliability. Nevertheless, faults can oc-

cur during operation. These may be caused by the following, e.g.:

Sensor

Process

Voltage supply

Signal processing

The first measure to be taken is to check the output signal. In many

cases, the causes can be determined this way and the faults rectified.

Should these measures not be successful, please call in urgent cases

the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since

we offer this service world-wide, the support is only available in the

English language. The service is free of charge, only the standard

telephone costs will be charged.

Connect a multimeter in the suitable measuring range according to

the wiring plan.

Error

Cause

Rectification

Current signal not

stable

Level fluctuations Adjust damping in the signal condi-

tioning instrument or process control

system

Current signal

missing

Wrong connection

to voltage supply

Check connection according to

chapter "Connection steps" and if

necessary, correct according to chap-

ter "Wiring plan"

No power supply Check cables for breaks; repair if nec-

essary

Operating voltage

too low or load re-

sistance too high

Check, adapt if necessary

Reaction when malfunc-

tions occur

Failure reasons

Fault rectification

24 hour service hotline

Checking the current

signal