10 Supplement

VEGABAR 66 • Foundation Fieldbus - climate compensated

36793-EN-130321

Material

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

9 … 32 V DC

Ʋ Ex ia instrument

9 … 24 V DC

Ʋ Ex d instrument

9 … 32 V DC

Operating voltage with illuminated display and adjustment module

Ʋ Non-Ex instrument

12 … 32 V DC

Ʋ Ex-ia instrument

12 … 24 V DC

Ex d instrument

12 … 32 V DC

Power supply by/max. number of sensors

Ʋ H1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures
Protection rating

Ʋ Transmitter

IP 68 (25 bar)

Ʋ Housing, standard

IP 66/IP 67

21)

Ʋ Aluminium and stainless housing

(optionally available)

IP 68 (1 bar)

22)

Ʋ External housing

IP 65

Overvoltage category

III

Protection class

II

Approvals
Instruments with approvals can have different technical data depending on the version.
For that reason the associated approval documents of these instruments have to be carefully

noted. They are part of the delivery or can be downloaded under www.vega.com via "VEGA Tools"

and "serial number search" as well as via "Downloads" and "Approvals".

10.2 Information on Foundation Fieldbus

Block diagram, measured value processing

The following illustration shows the Transducer Block and Function block in simplified form.

21)

Instruments with gauge pressure measuring ranges cannot detect the ambient pressure when submerged, e.g.

in water. This can lead to falsification of the measured value.

22)

Only with instruments with absolute pressure ranges.