VEGA VEGABAR 52 4 … 20 mA_HART - climate compensated User Manual

Page 50

Advertising
background image

50

10 Supplement

VEGABAR 52 • 4 … 20 mA/HART - climate compensated

36718-EN-130321

Ʋ Connection cable with IP 68 1 bar

version

PE

Weight approx.

0.8 … 8 kg (1.764 … 17.64 lbs), depending on process

fitting

Output variable
Output signal

4 … 20 mA/HART

HART output values according to HART standard 5.0

Ʋ Primary Value

Pressure

Ʋ Secondary Value

Temperature

9)

Signal resolution

1.6 µA

Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Max. output current

22 mA

Load

see load diagram under Power supply

Met NAMUR recommendation

NE 43

Dynamic behaviour output
Run-up time approx.

10 s

90 %

100 %

10 %

t

t

T

A

t

t

S

2

1

Fig. 27: Sudden change of the process variable. t

T

: dead time; t

A

: rise time; t

S

: jump response time

1 Process variable

2 Output signal

Dead time

≤ 150 ms

Rise time

≤ 100 ms (10 … 90 %)

Step response time

≤ 250 ms (ti: 0 s, 10 … 90 %)

Damping (63 % of the input variable)

0 … 999 s, adjustable

Additional output parameter - temperature
Processing is made via output signal HART multidrop, Profibus PA and Foundation Fieldbus
Range

-50 … +150 °C (-58 … +302 °F)

Resolution

1 °C (1.8 °F)

9)

Depending on the respective instrument and the process fitting, the temperature can deviate from the real

process temperature.