5

Setup

5

.1 Setup

After mounting a new electronics module and connecting to power
supply, the following data are loaded automatically from the meas. cell
electronics:

l

Instrument type

l

Serial number

l

Date of manufacture

l

Measuring range

l

Default

The instrument is generally ready for operation.

If you are using an electronics module without sensor serial number or
a "universal electronics module" from stock, you can load the order-
specific data of the sensor on site.

Information:
When loading on site, the corresponding file must first be downloaded
from the Internet (see "Load sensor data").

Load sensor data

Move via www.vega.com to the item "serial number search". The order
data of the sensor will be displayed after entering the serial number.

Below the order data you can find "Sensor data for Service DTM" as
XML file. Load this file with "Save target under" onto your PC and then
transfer the file via PACTware and the Service DTM to the sensor.

As a rule, settings already carried out on site with the previous
electronics module such as min./max. adjustment etc., however must
be repeated.

Tip:
Simply use the copy function of the adjustment module PLICSCOM or
the adjustment software PACTware.

Setup

Adjustment

Electronics module • VEGADIF 65