6 Setup
VEGASOURCE 31 •
38131-EN-130415
180°
1
3
6
3
4
5
2
2
1
Fig. 24: Switching the radiation on - VEGASOURCE 31 version B
1 Padlock
2 Locking screw
3 Fixing pin
4 Safety catch
5 Holding pin
6 Swivel bracket
Initial situation: Source holder is in position OFF
•
Remove padlock (1)
•
Loosen locking screw (2)
•
Pull out fixing pin (3)
Warning:
Do not remove the safety catch (4). Do not remove the seal. Do not
push the sealed pin (5). Turning the insert above the pin (5) leads to
the removal position of the source insert.
•
Turn stirrup (6) 180° anticlockwise
Note:
The actual switch position is shown by the visible label (ON or OFF).
The other label is always covered by the stirrup (6).
•
Let the fixing pin (3) snap in in position ON. Check if it snaps in
correctly
•
Fasten the padlock (1) in the specified position on the securing bar
(4) until it is required again
•
Tighten locking screw (2)
Radiation ON
The label EIN - ON is visible. The marking arrow points to EIN - ON.
Radiation OFF
The label AUS - OFF is visible. The marking arrow points to AUS -
OFF.
Indication of the switch-
ing status