VEGA VEGABAR 52 4 … 20 mA User Manual

Page 46

Advertising
background image

46

10 Supplement

VEGABAR 52 • 4 … 20 mA

36716-EN-130321

0,15 %

-0,15 %

-40°C

-20°C

20°C

40°C

60°C

80°C

Fig. 19: Thermal change, current output

Long-term stability (according to DIN 16086 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA under refer-

ence conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal meas-

uring range/set span.
Long-term drift of the zero signal:

Ʋ For one year

< 0.05 % x TD

Ʋ For five years

< 0.1 % x TD

Ʋ For ten years

< 0.2 % x TD

Ambient conditions
Ambient, storage and transport temperature

Ʋ Standard version

-40 … +80 °C (-40 … +176 °F)

Ʋ Versions IP 66/IP 68 (1 bar) and IP 68

(25 bar), connection cable PUR

-20 … +80 °C (-4 … +176 °F)

Ʋ Version IP 66/IP 68 (1 bar), connec-

tion cable PE

-20 … +60 °C (-4 … +140 °F)

Process conditions
The specifications of the pressure stage and product temperature are used as an overview. The

specifications on the type label are applicable.
Pressure stage, process fitting

Ʋ Thread 316L, depending on connec-

tion

PN 10, PN 60, PN 160

Ʋ Thread Aluminium

PN 25

Ʋ Thread PVDF

PN 10

Ʋ Hygienic fittings 316L, depending on

connection

PN 6, PN 10, PN 25, PN 40 (PN 40 only with DRD and

DIN 11851)

Ʋ Flange 316L

PN 16, PN 40,150 lbs, 300 lbs, 600 lbs

Ʋ Flange with extension 316L

without PN specification, PN 16, PN 40 or 150 lbs,

300 lbs, 600 lbs

Ʋ Flange flattened on both sides 316L

PN 10

Ʋ Flange PVDF

PN 16