VEGA VEGADIS 81 User Manual

Page 26

Advertising
background image

26

9 Supplement

VEGADIS 81 •

43814-EN-130502

Ʋ Inspection window

Polyester foil

Adjustment circuit
Connection to

the respective sensor

Data transmission

digital (I²C-Bus)

Configuration connection cable

4-wire, screened

Cable length max.

Ʋ Radar sensors with software ≥ 4.0.0,

hardware ≥ 2.0.0

50 m

Ʋ TDR sensors with software ≥ 1.0.0,

hardware ≥ 1.0.0

50 m

Ʋ Radiation-based sensors

50 m

Ʋ Radar and ultrasonic sensors, capaci-

tive sensors, pressure transmitters

with software ≤ 3.99, hardware < 2.0.0

25 m

Sensors with signal output Profibus PA or

Foundation Fieldbus

25 m

Cable length depending on the sensor version
Voltage supply heating
Operating voltage

Ʋ Type

Reliably separated extra low voltage according to VDE

0106 part 10

Ʋ Voltage height

24 V DC +5 %

Interpolation protection

Integrated

Power consumption max.

1.7 W

Switching reaction

Ʋ Heating on

< -5 °C (23°F )

Ʋ Heating off

> 0 °C (32 °F )

Electrical protective measures
Protection rating

Ʋ Housing plastic

IP 66/IP 67

Ʋ Housing Aluminium, stainless steel

IP 66/IP 68 (0.2 bar)

Approvals
Instruments with approvals can have different technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted.

They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as

well as under "Downloads" and "Approvals".

Advertising