6 Set up with the display and adjustment module

VEGAFLEX 81 • 4 … 20 mA/HART - four-wire

42280-EN-130314

In this menu item, you can define the medium type (product).
This menu item is only available if you have selected level measure-

ment under the menu item "Application".

You can choose between the following medium types:

Dielectric con-

stant

Medium type

Examples

> 10

Water-based liq-

uids

Acids, alcalis, water

3 … 10

Chemical mix-

tures

Chlorobenzene, nitro lacquer, aniline,

isocyanate, chloroform

< 3

Hydrocarbons

Solvents, oils, liquid gas

This menu item is only available, if you have chosen interface meas-

urement under the menu item "Application". In this menu item you can

enter if there is a superimposed gas phase in your application.
Only set the function to "Yes", if the gas phase is permanently pre-

sent.

This menu item is only available if you have selected interface meas-

urement under the menu item "Application". In this menu item you can

choose the medium type of the upper medium.

You can enter the dielectric figure of the upper medium directly or

have the figure determined by the instrument. For this purpose you

have to enter the measured distance to the interface.

In this menu item you can enter the max. adjustment for the level. With

interface measurement this is the maximum total level.

Adjust the requested percentage value with [+] and save with [OK].

Setup - Application - Me-

dium, dielectric constant

Setup - Application - Gas

phase

Setup - Application - Di-

electric constant

Setup - Max. adjustment

Level