11 Supplement

VEGABAR 83 • 4 … 20 mA/HART

45034-EN-131010

Nominal range

Overload capacity, max.

pressure

Overload capacity, min. pres-

sure

Gauge pressure
0 … +0.15 psig

+225 psig

0 … +5 psig

+375 psig

-11.60 psig

0 … +15 psig

+525 psig

-14.5 psig

0 … +30 psig

+600 psig

-14.5 psig

0 … +150 psig

+1350 psig

-14.5 psig

0 … +300 psig

+1500 psig

-14.5 psig

-14.5 … 0 psig

+500 psig

-14.5 psig

-14.5 … +20 psig

+580 psig

-14.5 psig

-14.5 … +150 psig

+1480 psig

-14.5 psig

-14.5 … +300 psig

+1575 psig

-14.5 psig

-0.7 … +0.7 psig

+225 psig

-14.5 psig

-3 … +3 psig

+290 psi

-14.5 psig

-7 … +7 psig

+510 psig

-14.5 psig

Absolute pressure
0 … 1.5 psi

225 psi

0 psi

0 … 5 psi

435 psi

0 psi

0 … 15 psi

510 psi

0 psi

0 … 30 psi

725 psi

0 psi

0 … 150 psi

1300 psi

0 psi

0 … 300 psi

1900 psi

0 psi

Output variable
For details on the operating voltage see chapter "Voltage supply"
Output signal

4 … 20 mA/HART

Range of the output signal

3.8 … 20.5 mA/HART (default setting)

Fulfilled HART specification

7.3

Signal resolution

0.3 µA

Failure signal current output (adjustable) ≥ 21 mA, ≤ 3.6 mA, last valid measured value
Max. output current

21.5 mA

Starting current

≤ 10 mA for 5 ms after switching on, ≤ 3.6 mA

Load

Load resistor, see chapter "Voltage supply"

Damping (63 % of the input variable),

adjustable

0 … 999 s

HART output values according to HART 7 (default setting)

5)

Ʋ First HART value (PV)

Linear percentage value

Ʋ Second HART value (SV)

Physical unit of the application

5)

The output values can be assigned individually