15

3 Mounting on a chemical tanker (free radiating)

VEGAPULS 63 • Cargo tank

45562-EN-130626

the effect of false echoes. The direction of the polarisation plane is

marked on the process fitting of the instrument.

1

4

2

3

Fig. 24: Polarisation - Orientation in case of heating coils
1 Polarisation marking in front

2 Polarisation marking on the side

3 Radar beam

4 Heating coils

Direct the radar beam to a spot without bulkheads. If this is not possi-

ble, the bulkheads should be covered with angle plates. These ensure

that the radar signals are deflected to the side instead of reflected

straight back to the sensor.

Fig. 25: Bulkheads in the radar beam: Deflection of the radar beam by angle

plates

Bulkheads