VEGA VEGABAR 81 HART and accumulator pack - Quick setup guide User Manual

Page 10

Advertising
background image

10

4 Connecting to power supply

VEGABAR 81 • HART and accumulator pack

46300-EN-131125

2

4

3

5

0

5

1

6

2

7

3

8

4

9

modes

4 ON/OFF

3 ON/OFF 1h

1,2 OFF

Charge

Accu

5..9 auto cycle

1

2

+

( )

(-)

1

Fig. 5: Supply room
1 Internal connection of the socket for the battery charger

2 Mode switch

3 LED green, charging process

4 LED yellow, charging status

5 Socket for the battery charger

The mode switch enables the selection of the following modes:

0 = sensor off, LEDs show the accumulator status

1, 2 = sensor off, LEDs off

3 = sensor on for 1 hour after pressing a key (delivery status)

4 = sensor permanently on, switching on/off via button

5 = sensor is switched on every 30 min. for 3 min.

6 = sensor is switched on every hour for 3 min.

7 = sensor is switched on every 6 hours for 3 min.

8 = sensor is switched on every 12 hours for 3 min.

9 = sensor is switched on every 24 hours for 3 min.

The green LED characterizes the charging process:

LED flashes = Accumulator is charging

LED lights = accumulator is full, battery charger should be

unplugged (accumulator life time)

After pressing the key or changing the mode the yellow LED shows

the accumulator status for approximately 10 s as follows:

LED lights = accumulator is full

LED flashes = accumulator should be charged

LED off = accumulator is empty

Supply room

Advertising