background image

16

6 Supplement

VEGABAR 86 • HART and accumulator pack

46320-EN-131125

6 Supplement

6.1 Technical data

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry

Ʋ Cable gland

M20 x 1.5

Ʋ Cable diameter (options)

5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)

Ʋ Cable entry

½ NPT

Ʋ Blind plug

M20 x 1.5; ½ NPT

Ʋ Closing cap

M20 x 1.5; ½ NPT

Plug options

Ʋ Signal circuit

Plug M12 x 1, according to ISO 4400, Harting HAN,

7/8" FF

Ʋ Indication circuit

M12 x 1 plug

Wire cross-section (spring-loaded terminals)

Ʋ Massive wire, cord

0.2 … 2.5 mm² (AWG 24 … 14)

Ʋ Stranded wire with end sleeve

0.2 … 1.5 mm² (AWG 24 … 16)

External battery charger
Mains voltage

100 … 240 V AC

Output voltage

24 V DC

Output current (shortcircuit proof) max. 500 mA
Load current limitation

70 mA

DC plug (inside plus, outside minus)

2.1 mm

Integrated accumulator
Type

Lithium ions

Voltage

14.8 V

Accumulator capacity

4.7 Wh

Charging period from 0 % to 100 %

approx.

4 h

Operating time after 10 min. charging of

0 %

> 3 h

Operating time in mode 4 (sensor per-

manently on) with completely charged

accumulator

> 60 h

Temperature range

Ʋ Charge accumulator

0 … +45° C (+32 … +167 °F)

Ʋ Accumulator operation

-20 … +60° C (-4 … +140 °F)

Temperature derating accumulator capacity

Ʋ +25° C (+77 °F)

100 %