6 supplement, 1 technical data – VEGA VEGABAR 87 Profibus PA - Quick setup guide User Manual
Page 17
17
6 Supplement
VEGABAR 87 • Profibus PA
46328-EN-131125
6 Supplement
6.1 Technical data
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland
M20 x 1.5
Ʋ Cable diameter (options)
5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)
Ʋ Cable entry
½ NPT
Ʋ Blind plug
M20 x 1.5; ½ NPT
Ʋ Closing cap
M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit
Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF
Ʋ Indication circuit
M12 x 1 plug
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Voltage supply
Operating voltage
Ʋ Non-Ex, Ex-d instrument
9 … 32 V DC
Ʋ EEx-ia instrument - Power supply
FISCO model
9 … 17.5 V DC
Ʋ EEx-ia instrument - Power supply
ENTITY model
9 … 24 V DC
Number of sensors per DP/PA segment coupler, max.
Ʋ Non-Ex
32
Ʋ Ex
10