Model 20/20c specifications – Trek 20-20C High-Voltage Power Amplifier User Manual

Page 2

Advertising
background image

Measurement and Power Solutions

TM

www.trekinc.com

Features (cont.)

Dynamic Adjustment

Graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters

Current Limit/Trip

Switch selectable for either limit or trip.
Graduated 1-turn panel potentiometer is used
to adjust limit or trip level from 0 to ±20 mA

Out of Regulation
Status

Illuminates and a TTL low is provided when unit
fails to produce required HV output such as
during current limit or short circuit load
conditions

Trip Status

Illuminates and a TTL low is provided when the
high-voltage output is disabled due to the
output current exceeding the current trip level,
the detection of a high-voltage supply fault or
the removal of the top cover

Fault Status

A BNC provides a TTL low when the Model
20/20C is out of regulation for greater than
500 ms

Mechanical

Dimensions

279 mm H x 482 mm W 654 mm D
(11” H x 19” W x 25.75” D)

Weight

24.9 kg (55 lb)

HV Connector

Caton High Voltage Connector

BNC Connectors

Amplifier Input, Voltage Monitor, Current Monitor,
Remote High Voltage ON/OFF, Out of Regulation
Status, Fault/Trip Status

Operating Conditions

Temperature

0°C to 40°C (32°F to 104°F)

Relative Humidity

To 85%, noncondensing

Altitude

To 2000 meters (6561.68 ft.)

Electrical

Line Voltage

Factory Set for one of two ranges:
104 to 127 V AC or 180 to 250 V AC,
either at 48 to 63 Hz

AC Line Receptacle

Standard IEC 320 three-prong AC line
connector

Power Consumption

1000 VA, maximum

Supplied Accessories

Operators’ Manual

PN: 23177

HV Output Cable

PN: 43466

Line Cord, Spare
Fuses

PN: N5011; s
destination

elected per geographic

Optional Accessories

19” Rack Mount Kit

Model: 608RA (with EIA hole spacing)
Model: 608RAJ (with JIS hole spacing)

Model 20/20C Specifications

Performance

Output Voltage Range

0 to ±20 kV DC or peak AC

Output Current Range

0 to ±20 mA DC or peak AC

Input Voltage Range

0 to ±10 V DC or peak AC

Input Impedance

25 kW, nominal

DC Voltage Gain

2000 V/V

DC Voltage Gain Accuracy

Better than 0.1% of full scale

DC Offset Voltage

Less than ±2 V

Output Noise

Less than 1.5 V rms*

Slew Rate
(10% to 90%, typical)

Greater than 450 V/µs

Large Signal Bandwidth
(-3 dB)

DC to greater than 7.5 kHz

Large Signal Bandwidth
(1% distortion)

DC to greater than 3.75 kHz

Small Signal Bandwidth
(-3dB)

DC to greater than 20 kHz

Stability

Drift with Time

Less than 50 ppm/hr, noncumulative

Drift with Temp

Less than 100 ppm/°C

Voltage Monitor

Ratio

1/2000th of the high-voltage output

DC Accuracy

Better than 0.1% of full scale

DC Offset Voltage

Less than ±2 mV

Output Noise

Less than 10 mV rms*

Output Impedance

47 W

Current Monitor

Ratio

0.5 V/ mA

DC Accuracy

Better than 1% of full scale

Offset Voltage

Less than ±10 mV

Output Noise

Less than 30 mV rms*

Bandwidth (-3dB)

DC to greater than 10 kHz

Output Impedance

47 W

Features

High-Voltage On/Off

Local

Individual push-button switch

Remote (TTL
compatible input)

TTL high (or open) turns off high-voltage
output. TTL low turns on high-voltage
output.

Copyright ã 2013 TREK, INC. All specifications are subject to change. 1322/JRB

*Measured using the true rms feature of the HP Model 34401A digital multimeter

Advertising