Trek’s Model 875

Electrostatic Voltage Sensor is designed for in-line monitoring

of electrostatic charge build-up, which if left unchecked, can disrupt
manufacturing processes or cause product degradation and early life failure of
semi-conductors and other charge-sensitive components. Excellent for
monitoring static charge levels within manufacturing equipment such as
conveyors, handlers, and other tools, the 875 features an automatic calibration
technique to maintain high accuracy and speed over wide variations of spacing
between the non-contacting measurement probe and the surface under test.

Other features include a voltage measurement range of ± 500 V DC or peak AC,
accuracy of ±0.5%, speed of 25 ms, and low noise of 1% rms of full scale.
Buffered output voltage and current monitors (4 to 20mA) are provided for remote
monitoring and alarm purposes. The unit is powered by +24 V DC and is housed in
a standard DIN package to enable standardized mounting.

!

Measurement Range of

0 to ±500 V DC or peak AC

!

Accuracy is Independent of

Probe-to-Measured Surface
Spacing

!

Voltage Monitor with an

Accuracy Better than
±0.5% of full scale (F.S.)

!

4-20 mA Current Monitor

!

Speed of Response

< 25 ms

!

Enclosure Mounts on 35mm

DIN RAIL

!

TTL Digital Enable Input

!

TTL Fault Warning Flag

Output

!

compliant

!

!

!

!

!

!

!

!

PERFORMANCE

FEATURES

PROBE

GENERAL

At 4 mm ±1 mm probe-to-surface

Probe Dimensions

Voltage Monitor Output

distance

5.6 mm square x 50 mm L

A buffered low-voltage replica of the
measured voltage.

Aperture Orientation / Body Type

Measurement Range

Scale Factor

Side / Square.

0 to ±500 V DC or peak AC.

1 V / 50 V.

Probe Cable Length

Noise

Measurement Accuracy

3.0 ±0.1 meters.

Less than 1% rms of full scale.

At the Voltage Monitor Output

Better than ±0.5% of F.S.

Recommended Probe-to-Surface

Digital Enable

At the Current Monitor Output

Separation

Better than ±3% of F.S.

4 mm ±1 mm.

An optional probe holder fixture is
available - contact factory.

Linearly related to the measured
input voltage.

Fault Warning Output

Power Supply Voltage

A TTL output signal. A TTL HIGH

+24 V DC ±10%.

indicates normal operation of the

Power Supply Current

Model 875. A TTL LOW indicates

150 mA, maximum.

a fault condition such as: out of
range operation (circuit is

Speed of Response (10% to 90%)

Power ON Indicator

measuring a voltage greater than

Less than 25 ms for a 0 to ±500 V

A LED indicator illuminates when

±500 Volts), failure of the probe,

step change.

power is applied to the unit.

or circuit malfunction.

Current Monitor (4-20 mA)

Scale Factor (+4 mA to +20 mA
representing a -500 V to +500 V):

-500 V = 4 mA,

0 V = 12 mA,

+500 V = 20 mA.

An external control TTL signal.
A TTL HIGH (or open) will disable
all internal power supplies.
A TTL LOW will provide normal
ESVM operation.