5. Overview

Refer to the illustrations on page 2 of this manual.

1 retractable hook tip

5 IC tip

2 adjustment tool

6 tip protector

3 ground lead with crocodile clip

7 identification tags

4 ground bayonet

6. Operating instructions

• Connect the BNC connector to the oscillator (not incl.). Refer to the user manual of the

oscilloscope for the correct port and operation settings.

• Make sure the probe is properly calibrated. Refer to §6 Calibration below.

• When possible, disconnect power to the circuit under test and connect the crocodile clip

to ground or common (usually the chassis) or use the ground bayonet [4] (only fits

over the standard tip).

Note: Always use the shortest possible ground lead! Excessive ground lead length

introduces unwanted inductance and can alter the displayed signal, as well as reduces

the scope's effective bandwidth (ground lead acts like a low pass filter).

• Depending on the measured signal, use the standard tip, the retractable hook tip [1] or

the IC-tip [5]. When wires are close together, place the tip protector [6] over the

standard tip to avoid short circuit.

• When using multiple probes on the same circuit board, clip the identification tags [7]

over the probe cable, one at the side of the BNC connector and the other near the

probe, to easily identify the signals.

• When possible, switch off power to the circuit under test prior to removing the probe.

7. Calibration

• In order to obtain the correct division ratio with each oscilloscope, the attenuation

network needs to be adjusted. To compensate the probe to your oscilloscope, apply a

1kHz square wave to the probe tip, or connect to the cal socket on the oscilloscope to

display a few cycles of the waveform (refer to the illustrations on page 2 of this

manual).

• Adjust the trimmer located in the BNC connector for a flat-topped square wave.

8. Technical specifications

attenuation x100
bandwidth

DC to 60MHz

rise time

5.8ns

input resistance

100MΩ (input resistance scope = 1MΩ)

input capacitance

±6.5pF

working voltage

1200Vdc incl. peak AC

compensation range

10 to 50pF

cable length

1.3m

remarks

with insulated BNC connector

Use this device with original accessories only. Velleman nv cannot be held

responsible in the event of damage or injury resulted from (incorrect) use of this

device. For more info concerning this product and the latest version of this user

manual, please visit our website www.velleman.eu. The information in this

manual is subject to change without prior notice.
© COPYRIGHT NOTICE

This manual is copyrighted. The copyright to this manual is owned by Velleman

nv. All worldwide rights reserved. No part of this manual may be copied, reproduced,

translated or reduced to any electronic medium or otherwise without the prior written

consent of the copyright holder.