Velleman DVM52IT User Manual

Page 3

Advertising
background image

DVM52IT

VELLEMAN

3

5.1 General Specifications


Max. voltage between terminals and earth

1000V dc or ac

Measuring Method

Dual-slope integration A/D converter

Sampling Time

±0.4sec.

LCD Height

22mm

Max. Display

1999 points (3 ½ digits)

Polarity Indication

‘-‘ indicates negative polarity

Overrange Indication

‘1’ or ‘-1’ is displayed

Displayed Measuring Unit

Unit of electrical capacity

Power Supply

6 AA-batteries of 1.5V (alkaline batteries are recommended)

Power Consumption

±5mA (100Vdc, 750Vac, 200

, )

±30mA (200M

, 250V)

±50mA (200M

/500V)

±100mA (2000M

/1000V)

Battery-Low Indication

”-symbol is displayed

Operating Temperature

0 to 40°C (32 to 104°F)

Operating Humidity

< 85%RH

Storage Temperature

-10 to +50°C (10 to 122°F)

Dimensions

192 x 122 x 55mm

Weight

± 545g (with batteries)


5.2 Electrical Specifications


5.2.1. Insulation Resistance

Range

200M

/ 250V

200M

/ 500V

2000M

/ 1000V

Test Voltage

250Vdc ± 10%

500Vdc ± 10%

1000Vdc ± 10%

0 ~2000M

Measuring Ranges

0 ~200M

0 ~200M

0 ~1000M

1000 ~2000M

Accuracy

± 3% rdg ± 5 digits

± 5% rdg ± 5 digits

Output Voltage on Open Circuit

250V ± 10%

500V ± 10%

1000V ± 10%

Min. Output Voltage

225V at 0.25M

450V at 0.5M

900V at 1M

Test Current (approx.)

1mA at 0.25M

1mA at 0.5M

1mA at 1M

Output Short-Circuit Current

2.5mA


5.2.2. AC Voltage

Range

Resolution

Accuracy

700V

1V

±1.2% of rdg + 5 digits


Input Impedance

10M

Max. Input Voltage

700V rms AC or 1000V DC

Frequency Range

40 to 400Hz

Response

average, calibrated in rms of a sine wave

Advertising