Velleman DCM267N User Manual

Page 6

Advertising
background image

DCM267N

00 (04/11/2008)

6

Velleman

®

• Insert the side of the temperature probe adapter marked with “+” in the

"V Ω °C/°F" jack [10] and the other side (“-”) in the "COM" jack [9].

• Hold the probe head to the device under test for ±30s and read the value

in the LCD display.

Do not change the rotary function switch when the thermocouple is

connected to the meter.

7. Battery

• When the low battery indication appears, replace the internal battery.

Always disconnect test leads when replacing the battery. Do not us the

device without batteries installed.

• The battery cover is located on the back of the device and is closed with

one screw. Remove the screw and open the battery compartment.

• Remove the battery and insert a new 9V battery following the polarity as

indicated in the battery holder. Only use a new battery of the same type
and specifications.

• Reinstall the cover and secure it with the screw.

Remove the battery when the device will not be used for a long time to avoid
leakage.

Do not recharge non-rechargeable batteries and do not throw in fire as they

may explode.

WARNING: handle batteries with care, observe warnings on

battery casing. Dispose of batteries in accordance with local
regulations.

Keep batteries away from children.

8. Technical Specification

DC voltage

0.4/4/40/400/600V

basic accuracy

±(1.5% of rdg + 3 digits) / ±(0.8% of rdg + 3

digits) for 0.4V range / ±(2.0% of rdg + 3 digits)
for 600V range

input impedance

7.8Mohm

maximum input

600V

AC voltage

0.4/4/40/400/600V

basic accuracy

±(1.8% of rdg + 5 digits) / ±(0.8% of rdg + 20
digits) for 0.4V range / ±(2.5% of rdg + 5 digits)

for 600V range

input impedance

7.8Mohm

frequency range

50-60Hz

maximum input

600V

AC current

40/400/1000A (max. line voltage 240V)

basic accuracy

±(2.5% of rdg + 10 digits) for 40A range / ±(2.5%
of rdg + 5 digits) for 400A range / ±(3.0% of rdg

+ 4 digits) for 1000A range

resistance 400/4K/40K/400K/4M/40M
basic accuracy

±(1.0% of rdg + 4 digits) for 400ohm range /
±(1.5% of rdg + 2 digits) for 4K~400K range /

±(2.5% of rdg + 3 digits) for 4M range / ±(3.5%
of rdg + 5 digits) for 40M range

overload protection 15s max. 250V (all ranges)