Velleman DVM13MFC2 User Manual

Page 5

Advertising
background image

DVM13MFC2

VELLEMAN

- 5 -

a.

Example 1: Measuring an analogue handset

Connect the test cable with the input from channel B and set the FUNCTION key to the 1-step position. Select the

gate time (adjust manually from 100 msec to 10 sec). The LED display will read:

b.

Example 2: Measuring a 30Mhz intercom emission frequency

Set the FUNCTION to the 2-step position and select the gate time (adjust manually from 100 msec to 10 sec). The

LED display will read:

c.

Example 3: Measuring a self-oscillating frequency (e.g. from a telephone, intercom etc.)

Set the FUNCTION key to the 2-step position. Connect one end of a cable with 5pF capacitance to the red clamp;

use the other end to measure the frequency by touching the contact point.

6. Warnings

When measuring a high-voltage or strong RF signal, make sure that the cables are in series and have a large

capacitance to avoid damage.

Press the RESET button once to reset the device or turn off the device in case the device is not working

appropriately.

When there is no input signal, the display may not necessarily read “0”. This is absolutely normal and does not

affect the measurements or the accuracy.

Do not expose the device to extreme temperatures, humidity, dirt, dust etc. Do not open the device to avoid lethal

electroshocks.

A strong interference will reduce the sensitivity when measuring.

7. Technical Specifications

Channel A (0.01Hz ~ 50MHz)

Frequency Range

DC couple 0.01Hz to 100Hz

AC couple 100Hz to 50MHz

Sensitivity

AC: 100Hz ~ 50MHz < 80m Vrms

DC: 0.01Hz ~ 1Hz ≤ 500m Vp-p

1Hz ~ 100Hz ≤ 80m Vrms

Impedance

1 Mohms

Attenuator

X1, X20

Max. Safety Voltage

30V (DC/AC peak)

Channel B (50MHz ~ 2.4GHz)

Frequency Range

from 50MHz to 2.4GHz

Sensitivity

50MHz ~ 1.2GHz ≤ 50m Vrms

1.2GHz ~ 2.4GHz > 80m Vrms

Coupling

AC only

Impedance

50 ohms

Max. Safety Voltage

3V

1-step

900MHz

2-step

30MHz