3472b – digital calliper, Introduction, Specifications & functions – Velleman 3472B User Manual

Page 3: Maintenance instructions, Troubleshooting, Applications (see p. 6)

Advertising
background image

3472B

3

PEREL

3472B – DIGITAL CALLIPER

1. Introduction

To all residents of the European Union

Important environmental information about this product

This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the
environment.
Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialised company for
recycling.

This device should be returned to your distributor or to a local recycling service.
Respect the local environmental rules.

If in doubt, contact your local waste disposal authorities.

Thank you for buying the DIGITAL CALLIPER! Read this manual carefully before bringing the device into service.

2. Specifications & Functions

Measuring range

Resolution & Repeatability

Accuracy

0-150mm / 0-6"

0.01mm / 0.0005"

0.02mm / 0.0001"

Data output

Serial output for interface with host computer or printer

Maximum measuring speed:

1.5m/sec or 60"/sec

Display:

LCD, features minus sign "-", character in 6.35mm / 0.25" height for 5 digits, small digit "5"
and "IN" indication for inch measuring unit

Battery:

one 1.5V silver oxide battery for one year of continuous use.

Low-battery warning by flashing display.

Operating temperature:

0°C

+40°C

Storage temperature:

-20°C

+70°C

Operating humidity:

0 to 80%

1. ON / OFF key
2. METRIC / INCH change over key (true conversion)
3. Zero setting (resets any value to zero for reference)
4. Battery cover

3. Maintenance Instructions

1. Keep the device and the battery compartment clean and prevent liquids from entering the electronic part to avoid damaged

electronics.

2. Clean the device with petrol, lubricate the body with some clock oil. Acetone and alcohol should not be used.

4. Troubleshooting

Problem

Cause

Solution

The 5 digits change every second

Battery voltage is under 1.45V

Replace battery.

Displayed value doesn't change.

Circuitry malfunction

Remove the battery and put it back after 30 seconds.

LCD shows 000.00mm / 00.000in

One of the keys is short circuited

Remove the cover & adjust key spring position

Function keys do not work

Key spring is bent or deformed

Remove the cover & adjust key spring position

Calliper does not measure
accurately

Sensor is dirty

Remove the cover and slider assembly, clean the
sensor with pressure air (

51kg/cm²), and clean it

with a little petrol for dirt prevention.

No display on LCD screen

1) No battery contact.
2) Battery voltage is under 1.1V.

1) Restore battery contact.
2) Replace battery

5. Applications (see p. 6)

For more info concerning this product, please visit our website www.perel.eu.

The information in this manual is subject to change without prior notice.

Advertising