Powersolve PS60ME Series User Manual

Medical desk top power supplies, Features

Advertising
background image

Powersolve Electronics Ltd. 8A Arnhem Road, Newbury, RG14 5RU. England

Tel: 01635 521858 Fax: 01635 523771 Email: [email protected]

Medical Desk Top Power Supplies

PS60ME Series
30-60 Watt Switch Mode
Universal AC Input

INPUT
Input Voltage Range

90 - 264Vac

Frequency

47 - 63Hz

Inrush Current

80A max. @ 240Vac

Conducted EMI

CISPR/FCC Class B

Isolation - Input/Output

5,656VDC

Leakage Current

0.1mA max.

OUTPUT
Hold-up Time

8mS typ. @ 115Vac

Short Circuit Protection

Continuous (Auto Recovery)

Over Voltage Protection

Yes

ENVIRONMENTAL
Operating Temperature Range

0°C to +40°C

Storage Temperature Range

-20°C to +85°C

Cooling

Natural convection

GENERAL
Safety Approvals

EN60601-1

Dimensions

132.0 x 58.0 x 30.5 mm
5.197 x 2.283 x 1.201 inches

Weight

345g

AC Inlet

IEC320/C8

Electrical Specification

Typical at 25°C, nominal line and 75% load unless otherwise stated.

Features

Universal Input Range 90-264VAC

Continuous Short Circuit Protection

Over Voltage Protection

Meets EN60601-1 & EN55011 Class B

CEC Level IV Compliant (Output cable length ≤1800mm)

(PS60ME050 Non-compliant)
(PS60ME120: Output cable ≤1200mm, 16AWG)

Optional Output Connectors

www.powersolve.co.uk

Output Voltage and Current Ratings

MODEL

OUTPUT

OUTPUT

RIPPLE &

VOLTAGE

LINE

LOAD

% EFF.

VOLTAGE

CURRENT

NOISE

SETPOINT

REG.

REG.

PS60ME050

5V

6A

50mV pk-pk

±4%

±1%

±6%

71%

PS60ME120

12V

5A

120mV pk-pk

±2%

±1%

±5%

83%

PS60ME150

15V

4A

150mV pk-pk

±2%

±1%

±3%

84%

PS60ME180

18V

3.33A

180mV pk-pk

±2%

±1%

±2%

84%

PS60ME190

19V

3.15A

190mV pk-pk

±2%

±1%

±2%

85%

PS60ME240

24V

2.5A

240mV pk-pk

±2%

±1%

±2%

85%

NOTE:
1. Voltage set point at 60% full load.
2. Output noise and ripple measured with 10µF electrolytic & 0.1µF ceramic capacitors across output at 20MHz BW.
3. Line regulation is measured from 100Vac to 240Vac at full load.
4. Load regulation is measured from 60% to full load and from 60% to 20% load (60% ±40% load).