3B Scientific Electron Deflection Tube D User Manual

3b scientific® physics

Advertising
background image


3B SCIENTIFIC® PHYSICS



1

Electron-Beam Deflection Tube D 1000651


Instruction sheet

11/12 ALF

5

7

2

1

-

4

6

2
1

1
2

10 9 8 7 6 5 4 3 2

3

1 Fluorescent screen
2 Lower deflection plate
3 Boss with 4-mm plug for

connecting deflection plates

4 Electron gun
5 4-mm sockets for connecting

heater supply and cathode

6 4-mm plug for connecting

anode

7 Upper deflection plate

1. Safety instructions

Hot cathode tubes are thin-walled, highly

evacuated glass tubes. Treat them carefully as

there is a risk of implosion.

Do not subject the tube to mechanical stresses.

Do not subject the connection leads to any

tension.

The tube may only be used with tube holder

D (1008507).

If voltage or current is too high or the cathode is

at the wrong temperature, it can lead to the tube

becoming destroyed.

Do not exceed the stated operating parameters.

Only change circuit with power supply

equipment switched off.

Only exchange tubes with power supply

equipment switched off.

When the tube is in operation, the stock of the

tube may get hot.

If necessary, allow the tube to cool before

dismantling.

The compliance with the EC directive on elec-

tromagnetic compatibility is only guaranteed

when using the recommended power supplies.

2. Description

The electron-beam deflection tube is intended

for investigating the deflection of electron beams

in electrical and magnetic fields. It can be used

to estimate the specific charge of an electron
e

/m and to determine the electron velocity v.

The electron-beam deflection tube comprises an

electron gun which emits a narrow, focussed

ribbon of cathode rays within an evacuated,

clear glass bulb. A tungsten 'hairpin' filament hot

cathode is heated directly and the anode takes

the form of a cylinder. The deflection of rays can

be achieved electrostatically by means of a built-

in plate capacitor formed by the pair of deflec-

tion plates or magnetically with the help of the

Helmholtz coils D (1000644) magnetically. The

cathode rays are intercepted by a flat mica

sheet, one side of which is coated with a fluo-

rescent screen and the other side of which is

printed with a centimetre graticule so that the

path of the electrons can be easily traced. The

mica sheet is held at 15° to the axis of the tube

by the two deflecting plates.


Advertising