Geiger-Muller Tube 1000661

Instruction sheet

10/12 ALF


1 Window
2 Holder
3 Patch

cord

4 BNC

connector

1. Safety instructions

The window is fragile and can easily be dam-
aged. This will make the end-window counter
inoperable.

Do not touch the window.

The G-M tube can be damaged by self-
maintained gas discharges, if operated at too
high a voltage.

Do not exceed the operating voltage of 500
V in a sustained manner.


2. Description

The Geiger-Muller tube is a self-quenching
Halogen -trigger counter tube for registering
and X radiation.
The G-M tube comprises a counter tube with a

very thin mica window and a removeable holder
for mounting on the swivel arm of the X-ray ap-
paratus (1000657 or 1000660) and a plated
spring clip for alternative mounting. A BNC-
terminated coaxial cable connects the output of
the G-M tube to a counter.

2. Technical data

Dose rate range:

10

-3

to 10

2

mGy/h

Mass of the active
surface (mica):

2.0 to 3.0 mg/cm²

Operating voltage:

500 V

Output connector:

BNC

Cable length:

1 m

Dimensions:

57 x 22 mm diam.