3B Scientific pH Sensor User Manual

Page 5

Advertising
background image

2

4. Technical data

Measurement range:

pH 0 to pH 14

pH at zero voltage:

pH 7.00 ± 0.25

Sensor type:

Ag-AgCl combination elec-
trode, gel-filled, non-
refillable

Accuracy:

pH 0.05 in the range from
20°C to 25°C

Resolution: pH

0.01

Reaction time:

1 s for 95% of final value

5. Operation

Remove storage bottle from the electrode by
twisting off the lid, then pull the lid off the elec-
trode.

Rinse the lower end in the vicinity of the glass
bulb thoroughly with distilled water, shake off
remaining water and dry with filter paper.

Move air bubbles in the lower part of the elec-
trode chamber into the upper part by cautiously
shaking them downward.

Calibrate the electrode. For the procedure, see
section 6 “Calibration”.

Subsequently, rinse with distilled water, shake off
remaining water and dry with filter paper.

Conduct measurement.

After finalising the measurement, rinse the elec-
trode with distilled water and repeat the same
procedure as before the measurement.

Slide the lid on the electrode and tighten it to the
storage bottle.


6. Calibration

Connect the electrode to the sensor box, and
connect the sensor box to analog input A of the
3B NETlog

TM

unit, analog input B is suitable for

pH-measurement but not for calibration!

Wait for the “Probe Detect” function to detect the
sensor.

Press the (Date/Time

↵) key 4 times until “User

Calibration” appears in the display.

Start calibration by pressing the (Store

L) key.

Immerse the electrode (after rinsing) in the alka-
line “pH 9.00 buffer solution”.

Move the decimal points in the display of the (SET
1) with the (Channel

I) key towards the left or

with the (Date/Time

↵) key to the right, respec-

tively.

Increase the displayed value with the (Rate

K) key,

or decrease it with the (Store

L) key, respectively.

Set the displayed value to +9.000e+00 and press
(Date/Time

↵).

Wait for the correction of the measured value,
press (

J

ACCEPT) and validate with the

(Date/Time

↵) key.

Remove the electrode from the “pH 9.00 buffer
solution”, rinse and dry thoroughly, and immerse
in the “pH 4.00 storage solution”.

Repeat the procedure described above for (SET 2).

Wait for the “Probe Detect” function to detect the
sensor then conduct the desired measurements.

7. Cleaning the electrode

After prolonged periods of use, clean the elec-
trode for 1 to 2 minutes in dilute hydrochloric
acid HCl (0.1 mol), then rinse with distilled water
and rejuvenate in potassium chloride KCl (4 mol).

8. Use in experiments

Measurement of common household acids and alkalis

Acid-base titrations

Observation of pH changes in chemical reactions

Testing the influence of photosynthesis on water qual-
ity in aquaria

Examining the influence of acid rain and its neutrali-
sation

Water quality of rivers and lakes

9. Sample experiment

9.1 Qualitative determination of the titration

curve of vinegar

Required equipment:

1 3B NETlog

TM

U11300

1 pH Sensor

U11350

1 Measuring cylinder, 100 ml

U14205

Household vinegar pH 2.00

Highly dilute sodium hydroxide pH < 14.00

Advertising