3B Scientific Electrometer Box User Manual

3b scientific® physics

Advertising
background image


3B SCIENTIFIC® PHYSICS



1

Electrometer Box U11375

Instruction sheet

01/09 Hh

1. Safety instructions

The electrometer box has an extremely high
resistance at its voltage input, and can be damaged if
an excessive voltage is applied.

Do not exceed the maximum input voltage of ±8
V!

If necessary connect voltage-limiting components
to the input!


2. Description

The electrometer box is an impedance converter with
an extremely high input resistance for measuring very
small charges or currents.

It can be used in conjunction with the 3B NETlog

TM

unit (U11300) for single manual measurements or
with 3B NETlab

TM

(U11310) for recording and analysing

measurements.

The electrometer box is recognised automatically by
the 3B NETlog

TM

unit..


3. Equipment supplied

1 Electrometer box

1 miniDIN 8-pin connecting lead, 600 mm long

1 Instruction sheet for U11375

4. Technical data

Input resistance:

> 10

11

ohms

Input capacitance:

< 50 pF

Measurement error:

< 1.5%

Overvoltage tolerance for non-hazardous
contact voltages:

1 kV (for low-
resistance

sources)

10 kV (for high-resistance
sources)

Connectors:

4-mm safety sockets


5. Operation

Connect the electrometer box to an analog input
of the 3B NETlog

TM

unit (input A or B).

Short-circuit the input and calibrate the offset so
that the output reading is zero.

Carry out the chosen experiment without delay so
that there is not time for stray charge to collect
on the input terminals.

Before starting a new experiment short-circuit the
input again.

Advertising