Ocean Optics R200 User Manual

R200 reflection probes, Operation, Specifications

Advertising
background image

R200 Reflection Probes

R200 R

EFLECTION

P

ROBES

couple to Ocean Optics miniature fiber optic spectrometers and light sources to create

small-footprint optical-sensing systems for fluorescence and reflection measurements. Ocean Optics offers several
variations on the Reflection Probe:

!

The R200-7 R

EFLECTION

P

ROBE

consists of a bundle of 7 optical fibers -- 6 illumination fibers around 1 read

fiber -- each of which is 200

µ

m in diameter. A 3.0" x 0.25" stainless steel ferrule houses the fiber bundle.

Other options of this standard probe assembly include the R400-7, which has a bundle of optical fibers 400

µ

m in diameter, and the RP200-7, which has a 3.0" x 0.25" plastic ferrule to house the fiber bundle. The

RP200-7 is useful where a stainless steel ferrule may not be suitable, such as some applications involving
corrosive samples.

!

For reflection experiments across the UV-VIS-Shortwave NIR (200-1100 nm), there is the R200-MIXED.
The R200-MIXED consists of fourteen 200

µ

m-diameter optical fibers -- 6 UV/VIS and 6 VIS/NIR

illumination fibers, plus 1 UV/VIS fiber and 1 VIS/NIR read fiber. In addition, the R200-MIXED has a 3.0" x
0.25" stainless steel ferrule to house its fiber bundles, and couples easily to a dual-channel spectrometer in
which each channel is set for a different wavelength range.

!

Also available is the R200-REF, which consists of an R200-7 Reflection Probe and an additional fiber optic
to monitor an illumination source such as our LS-1 Tungsten Halogen Light Source. The R200-REF is useful
when a reflection experiment either does not allow the user to take frequent reference scans or includes an
illumination source with an unstable spectral output.

!

One other option is the R200-ANGLE R

EFLECTION

P

ROBE

, which has a bundle of seven 200-

µ

m fibers -- 6

illumination fibers around 1 read fiber -- and a 3.0" x 0.25" stainless steel ferrule with a 45° window. This angled
window removes the effects of specular reflection when the probe is immersed in dense liquids and powders.

Operation

The reflection probe consists of a 6-fiber leg (the illumination leg) that should be coupled to the light source, and a
single-fiber leg (the read leg) that attaches to the spectrometer.
1.

Inspect the ends of the fiber legs. The hole in the SMA connector is noticeably larger in the illumination leg
than the hole in the read leg. Also, the read leg should have colored bands.

2.

Attach the illumination leg to the light source. Attach the read leg to the spectrometer.

3.

Using the RPH-1 Reflection Probe Holder or some other holding device, point the probe at the surface to
be measured.

4.

The distance from the probe tip to the sample directly affects the signal. For quantitative results, the distance
and angle must be held constant.

Specifications

Operating Instructions: R200

Fiber core diameter:

200

µ

m or 400

µ

m

Fiber core/cladding:

silica

Fiber bundle:

6 illumination fibers around 1 read fiber

Numerical aperture:

0.22

Optimization:

UV/VIS (200-750 nm) and VIS/NIR (450-1000 nm)

Ferrule:

Stainless steel or plastic

Ferrule dimensions:

3.0" x 0.25"

Terminations:

SMA 905

Sheathing:

Blue PVC with Kevlar reinforcement

Temperature range:

-20° C to 80° C

Probe assembly length:

2 meters (breakout is halfway); custom probes are also available

Advertising