Next, store a dark spectrum. Either switch off the lamp, or
place an opaque object in the cuvette holder to block the
light path. Then click the Store Dark button.

Note:

The lamp must be ON before you click the Finish button.

Fig. 20

Click Finish.
A dialog box appears giving you three choices. See
Fig. 21.

Beer-Lambert Law

Calibrate from solutions of known concentration

Load calibration from file


Notice that an Absorbance graph is running in the
chart window. See Fig.21.

Fig.21

BEER-LAMBERT LAW

Click the radio button for Beer-Lambert Law.

Enter the data. The value of ε will either come from a
reference text or from the experiments with known con-
centrations described in the next section.
Click Finish to see the concentration reading. Now you
can use any concentration in a cuvette and get an in-
stant concentration reading by clicking Single Update or
Continuous Update.

To run another concentration measurement use
File>New>New Concentration Measurement

, but use

Active Processing

because you have an active

absorbance graph already running.

Fig. 22

Fig. 23