USB2000 Component Table

Ocean Optics permanently secures all components in the USB2000 at the time of manufacture. Only Ocean
Optics Technicians can replace interchangeable components, where noted.

Item Name

Description

1 SMA

Connector

The SMA Connector secures the input fiber to the spectrometer. Light from the
input fiber enters the optical bench through this connector.

2 Slit*

The Slit is a dark piece of material containing a rectangular aperture, which is
mounted directly behind the SMA Connector. The size of the aperture regulates the
amount of light that enters the optical bench and controls spectral resolution.

You can also use the USB2000 without a Slit. In this configuration, the diameter of
the fiber connected to the USB2000 determines the size of the entrance aperture.

Only Ocean Optics technicians can change the Slit.

3 Filter*

The Filter is a device that restricts optical radiation to pre-determined wavelength
regions. Light passes through the Filter before entering the optical bench. Both
bandpass and longpass filters are available to restrict radiation to certain
wavelength regions.

Only Ocean Optics technicians can change the Filter.

4 Collimating

Mirror

The Collimating Mirror focuses light entering the optical bench towards the Grating
of the spectrometer.

Light enters the spectrometer, passes through the SMA Connector, Slit, and Filter,
and then reflects off the Collimating Mirror onto the Grating.

5 Grating*

The Grating diffracts light from the Collimating Mirror and directs the diffracted light
onto the Focusing Mirror. Gratings are available in different groove densities,
allowing you to specify wavelength coverage and resolution in the spectrometer.

Only Ocean Optics technicians can change the Grating.

6 Focusing

Mirror

The Focusing Mirror receives light reflected from the Grating and focuses the light
onto the CCD Detector or L2 Detector Collection Lens (depending on the
spectrometer configuration).

7

L2 Detector
Collection Lens*

The L2 Detector Collection Lens (optional) attaches to the CCD Detector. It
focuses light from a tall slit onto the shorter CCD Detector elements.

The L2 Detector Collection Lens should be used with large diameter slits or in
applications with low light levels. It also improves efficiency by reducing the effects
of stray light.

Only Ocean Optics technicians can add or remove the L2 Detection Collection
Lens
.

8

CCD Detector
(UV or VIS)

The CCD Detector collects the light received from the Focusing Mirror or L2
Detector Collection Lens and converts the optical signal to a digital signal.

Each pixel on the CCD Detector responds to the wavelength of light that strikes it,
creating a digital response. The spectrometer then transmits the digital signal to the
OOIBase32 application.