OOINLCorrect Loading Non-Linearity Correction Coefficients Instructions

Note

OOINLCorrect only works on Windows operating systems.

Procedure

The following procedure uses the USB2000 Spectrometer as an example.

To correct for non-linearity using OOINLCorrect software, do the following:

1. Install OOINLCorrect software from the Software and Technical Resources CD or download

it from our Ocean Optics website at

http://www.oceanoptics.com/Technical/softwaredownloads.asp

. You will need to get the

password from an Ocean Optics Application Scientist to run this software.

2. Run the OOINLCorrect software by selecting Start | Programs | Ocean Optics |

OOINLCorrect | OOINLCorrect.

3. Choose Spectrometer | Hardware Configuration from the menu to set system parameters

for your spectrometer model as follows:

Parameter Select

Spectrometer type

Your spectrometer model. For USB4000 and HR4000, select S4000.

A/D Type

The appropriate A/D Converter for your spectrometer.

USB Serial Number The serial number corresponding to your spectrometer (if you have

more than one spectrometer connected to you system)

4. Click OK.

5. Set the software parameters on the right side of the screen as follows:

Parameter Setting

Channel Master

Polynomial Fit

7

th

Order

Integ. Period

300

Average 8

Boxcar smooth

10

Low Integration

10

High Integration

300

Integration Step

10

Saturation val

Enter the appropriate saturation value for your spectrometer:

S2000/USB2000/HR2000: 3800
HR2000+/HR4000: 15000
USB4000/QE65000: 63000

2

000-20000-200-01-0507