Product Specifications

Materials:

Biocompatible Black Delrin body bonded with a
Non-leaching solvent acrylic adhesive
to a 50 micron UV transparent film.

Connectors:

Hose barb connector (CFV-PUMP) is Polysulfone seated in acrylic
Pipette interface (CFV-PIP) is acrylic

Contained volume, 50 uL
Pathlength 0.75 mm
Dead Volume in channels: 1 uL for the inlet channel, 1 uL for outlet
channel

FluoroVette Overall dimensions: 4.7 mm tall, 4.3 mm wide, and 1 mm thick.
Adaptor dimensions: 15 mm from base to the center of light path, 1 cm x 1 cm
outer dimension. Insert retaining ring with or without a lens to reduce light scatter
NOTE: The adaptor has a separate retaining ring to allow insertion of a plano-
convex lens. Suggested lens; Edmund Optics: www.edmundoptics.com

LENS PCX 5 X 4.6 MGF2

COM

Dia.

5.0

E.F.L.

4.6

Type

PCX

Comment

LAKN14

Part #

NT32-403

Price

$19.00

Solvent compatibility:

Ethanol ( up to 99.5%)
Isopropyl alcohol ( up to 91% in water)
Aqueous buffers with or without detergents

DO NOT FILL with Acetone or Aliphatic solvents such as Hexane. Wipe
the outside with lens tissue to avoid scratching the Optical Film.

Directions for use:

The FluoroVette has an inlet side, which allows the flowcell to fill from the bottom to
minimize air bubble formation. The inlet side is shown on the left, the outlet side is
shown on the right in Figure 1.
NOTE Insert the FluoroVette

TM

into the adaptor with the inlet side facing

the light source of the fluorometer. Two notches at the top of the adaptor
indicate the light path as shown in Figure 2

Figure 1

Figure 2

FluoroVette

TM

with Hose Barb Connections

Shown below with the Adaptor assembly.





Inlet

Outlet

Side

Side



Light Source

FluoroVette

TM

with Pipette Interface


ALine, Inc. PO Box 1068, Redondo Beach, CA 90278

830 Douglas Ave., Dunedin, FL 34698

Customer

support:

877-707-8575

727-733-2447

Advertising