S1024dw miniature fiber optic spectrometer – Ocean Optics External Triggering Options User Manual

Page 13

Advertising
background image

External Triggering Instructions

200-00000-000-01-201401

13

S1024DW Miniature Fiber Optic Spectrometer

Integration Time

SAD500

(500 kHz Master Clock

Freq)

Integration Time

Equation

Jumper Block 4 (JP4)

Pins 1-2

Pins 3-4

5 ms

1000/F

m

open

open

60 ms

12000/F

m

open

shorted

300 ms

60000/ F

m

shorted

open

655 ms

131070/ F

m

shorted

shorted

F

m

= the master clock frequency in Hz (i.e., 200 kHz for the SAD500); shorted = connected; open = disconnected

Procedure

1. Note the configuration of the pins in the Jumper Block 4 column of the chart.

2. Remove the spectrometer from its housing. Do not tamper with the optical bench.

(If you have more than one channel in your system, you may have to disconnect the
channels from one another. The master spectrometer is always on the bottom of a
multiple channel system.)

3. Locate Jumper Block 4 (JP4) in the center of the green circuit board, near the

optical bench. (See the figure to the right.) Jumper Block 4 consists of the first four
pins, which are numbered 1, 2, 3, and 4.

4. Using jumpers, configure the pins to match the integration time you selected.

For example, if you have an S2000 and an ADC1000 A/D Converter, you select
13.3 milliseconds as the integration time. To enable this, place a jumper over pins 3 and 4 to short
the pins, and leave pins 1 and 2 open.

5. Click on the Data Acquisition icon or select Spectrum | Configure Data Acquisition from the

OOIBase32 menu bar, then set other acquisition parameters (averaging, boxcar smoothing, etc.)
in the software.

6. Select the External Trigger page from the Configure Data Acquisition screen, and then select

Hardware Trigger.