Advanced instruments inc, Sampling, Standby – Analytical Industries GPR-2800 AIS ATEX Oxygen Analyzer User Manual

Page 30

Advertising
background image

Advanced Instruments Inc.



30

9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to

reach to the desired digit of the OUTPUT SPAN OFFSET value.

10. Press the ARROW keys to enter the OUTPUT SPAN OFFSET value.
11. Repeat steps 9 and 10 until the complete OUTPUT SPAN OFFSET value has been entered.

12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.

13. The system returns to the SAMPLING mode.

Sampling


GPR-2800AIS Oxygen Transmitter requires positive pressure to flow the sample gas by the sensor to measure the oxygen
concentration in a sample gas. If not available see Pressure & Flow section.

Note:
Prematurely initiating the ZERO CALIBRATION procedure can cause the analyzer to display a negative reading in both
the ZERO and SAMPLE modes. Prematurely initiating the SPAN CALIBRATION procedure can cause erroneously high offsets and
inaccurate readings.

Procedure:
1. Following calibration the transmitter returns to the SAMPLE mode after 30 seconds.
2. Select the desired sampling mode - auto or if manual, the range that provides maximum resolution – as described above.
3. Use metal tubing to transport the sample gas to the transmitter.
4. The main consideration is to eliminate air leaks which can affect oxygen measurements above or below the 20.9% oxygen

concentration in ambient air - ensure the sample gas tubing connections fit tightly into the 1/8” male NPT to tube adapter,
and, the NPT end is taped and securely tightened into the mating male quick disconnect fittings which mate with the
female fittings on the transmitter

5. Assure there are no restrictions in the sample line.
6. For sample gases under positive pressure the user must provide a means of controlling the inlet pressure between 5-30

psig and the flow of the sample gas between 1-5 SCFH, a flow rate of 2 SCHF is recommended

7. For sample gases under atmospheric or slightly negative pressure an optional sampling pump is recommended to draw the

sample into the transmitter. Generally, no pressure regulation or flow control device is involved.

8. Caution: If the transmitter is equipped with an optional sampling pump and is intended for use in both positive and

atmospheric/slightly negative pressure applications where a flow meter valve is involved – ensure the valve is completely
open when operating the sampling pump. Refer to the Pressure & Flow section above.

9. Assure the sample is adequately vented for optimum response and recovery – and safety.
10. Allow the oxygen reading to stabilize for approximately 10 minutes at each sample point.

To avoid erroneous oxygen readings and damaging the sensor:
¾ Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is flowing to the

sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).

¾ Assure there are no restrictions in the sample or vent lines
¾ Avoid drawing a vacuum that exceeds 14” of water column pressure – unless done gradually
¾ Avoid excessive flow rates above 5 SCFH which generate backpressure on the sensor.
¾ Avoid sudden releases of backpressure that can severely damage the sensor.
¾ Avoid the collection of particulates, liquids or condensation collect on the sensor that could block the diffusion of oxygen

into the sensor.

¾ If the transmitter is equipped with an optional integral sampling pump (positioned downstream of the sensor) and a flow

control metering valve (positioned upstream of the sensor), completely open the flow control metering valve to avoid
drawing a vacuum on the sensor and placing an undue burden on the pump.

Standby


The transmitter has no special storage requirements.
The sensor should remain connected during storage periods.
Store the transmitter with the power OFF.
If storing for an extended period of time protect the analyzer from dust, heat and moisture.

Advertising