Advanced instruments inc – Analytical Industries GPR-1535 GB Trace PPM Oxygen Transmitter User Manual

Page 23

Advertising
background image

Advanced Instruments Inc.



23

Symptom

Possible Cause

Recommended Action

Erratic O

2

reading

or
No O

2

reading







Test sensor independent from transmitter







Change in sample pressure


Dirty electrical contacts in upper section of
sensor housing


Corroded solder joints on sensor PCB from
corrosive sample or electrolyte leakage from
sensor

Corroded spring loaded contact in upper
section of sensor housing from liquid in
sample or electrolyte leakage from sensor



Liquid covering sensing area



Improper sensor selection


Presence of interference gases

Presence of sulfur gases

Unauthorized maintenance

Sensor nearing end of life

Remove sensor from housing. Using a volt-meter
set to uA output; apply the (+) lead to the outer
ring of the sensor PCB and the (-) lead to the
center circle to obtain the sensor’s output in air.
Contact factory with result.
Sensors without PCB use mV setting.

Calibrate the transmitter (calibrate at pressure
and temperature of sample)

Clean contacts with alcohol (minimize exposure
time of MS sensor to ambient air to extent
possible)

Replace sensor and return sensor to the factory
for warranty determination


Upper section of sensor housing: Clean contacts
with alcohol, flow sample or zero gas for 2-3
hours to flush sample system and sensor housing
Sensor: Replace if leaking and return it to the
factory for warranty determination

Wipe with alcohol and lint free towel or flow
sample or zero gas for 2-3 hours to flush

Replace GPR/PSR sensor with XLT sensor when
CO

2

or acid gases are present


Consult factory

Replace sensor and install scrubber

Replace sensor, obtain authorized service

Replace sensor


Erratic O

2

reading

or
Negative O

2

reading

or
No O

2

reading

possibly
accompanied by
electrolyte leakage


Pressurizing the sensor by flowing gas to the
sensor with the vent restricted or SHUT OFF
valve which places a vacuum on the sensor in
excess 4” of water column, something which is
strongly discouraged. The

front sensing

membrane is .000625 thick, heat sealed to the
sensor body and subject to tearing when
vacuum is suddenly applied.

A premature adjustment of the ZERO OFFSET
potentiometer is a common problem


Zero the transmitter. If not successful replace the
sensor

Avoid drawing a vacuum on the sensor





From MAIN MENU select DEFAULT ZERO