Analytical Industries GPR-2500 Series Oxygen Transmitter User Manual

Page 20

Advertising
background image



20

Establishing Power to the Electronics:
Once the two wires of the shielded cable are properly connected to the
terminals inside the junction box as described above, connect the other

end of the two wires to a suitable 12-36V DC power supply with

negative ground such as a PLC, DCS, etc.
The digital display responds instantaneously. When power is applied,

the transmitter performs several diagnostic system status checks termed

“START-UP TEST” as illustrated below:

START-UP TEST


ELECTRONICS – PASS
LOOP POWER – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS

REV. 1.61


Note: The transmitter display defaults to the sampling mode when 30

seconds elapses without user interface.

3.3

%

AUTO SAMPLING

10% RANGE

24.5 C 100 KPA


Menu Navigation:
The four (4) pushbuttons located on the front of the transmitter operate the micro-processor:
1. green ENTER (select)
2. yellow UP ARROW
3. yellow DOWN ARROW
4. blue MENU (escape)

Main Menu:
Access the MAIN MENU by pressing the MENU key: