CALIBRATION will appear and the analyzer will return to Sample mode without completing the Zero
calibration.

Both the Zero Calibrate and Span Calibrate functions result in the following displays:

PASSED CALIBRATION

OR



FAILED CALIBRATION

Default Zero

The feature will eliminate any previous zero calibration adjustment and display the actual signal output of the
sensor at a specified oxygen concentration. For example, assuming a zero gas is introduced, the display
above 0.00% will reflect an actual zero offset. This feature allows the user to ensure that the accumulative
zero offset never exceeds 50% of the lowest range limit. To perform Default Zero,

Access the MAIN MENU by pressing the MENU key.

Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.

Press the ENTER key to select the highlighted menu option.

The following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS

>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO

Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.

Press the ENTER key to select the highlighted menu option.

The following display appears and after 3 seconds the system returns to the SAMPLING mode:


FACTORY

DEFAULTS

SET

3.3 PPM

AUTO SAMPLING
10 PPM RANGE

76 F 100 KPA