Advanced instruments inc – Analytical Industries GPR-2000 ATEX Portable Oxygen Analyzer User Manual

Page 24

Advertising
background image

Advanced Instruments Inc.



24

MAIN MENU

AUTO SAMPLE
MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS

BYPASS ALARMS



>>>

CALIBRATION

SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO


4. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.

5. Press the ENTER key to select the highlighted menu option.

6. The following display appears

100.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN




7. Compute the adjustment value as described in Appendix B or consult the factory. The true adjustment value must be

determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.

099.0

OUTPUT SPAN OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE

MENU TO RETURN




8. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to

reach to the desired digit of the OUTPUT SPAN OFFSET value.

9. Press the ARROW keys to enter the OUTPUT SPAN OFFSET value.
10. Repeat until the complete OUTPUT SPAN OFFSET value has been entered.
11. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
12. The system returns to the SAMPLING mode.

Advertising
This manual is related to the following products: