Step 5 wire relay outputs – YSI 5200A User Manual

Page 24

Advertising
background image

YSI 5200A
46

YSI 5200A
47

Complete other wiring to 5200A.

5.

Close front panel -

6.

page 40.

Test DC power.

7.

Test DC Power

Connect DC power cable to power source. The enabled sensor values will display

on the Run Screen after the initial YSI splash screen. Verify DC power at the Menus

→ System →Version menu - page 125.

See Power Fail Menu - page 117 for DC low battery trigger monitor and alarm

information.

Note:

The 5200A will not power up if DC power supply wires are terminated

-

incorrectly.

Step 5 Wire Relay Outputs

The 5200A has four (4) on board non-fused relays rated at 240VAC@5AMPs

(110VAC@10AMPs) switching capacity. Relays energize and de-energize peripheral

control, alarm, and/or timer devices. Terminate relay device wires to Normally

Open (NO), Normally Closed (NC), and Common (C) sockets as required for the

application. (See 5200A I/O board wiring diagram on inside of front panel for iden-

tification of relay number and pin position.) Relays 1-4 are terminated at location

C on I/O board - figure 3.19.

CAUTION

:

Do not overload relay(s). Check manufacturer specifications for

inductive current/amp at power up. Use a contactor if necessary - figure 3.20.

Directions to Wire Relay Outputs

WARNING: Disconnect external power to the unit before wiring. If any of

the relay contacts are connected to a “hazardous-live” circuit, ALL contacts on

ALL relays must be considered “hazardous-live” and appropriately protected

from user contact outside the 5200A enclosure.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant

d’effectuer le câblage. Si un contact de relai quelconque est connecté à un

circuit sous tension dangereuse, TOUS les contacts de TOUS les relais doivent

être considérés comme étant sous tension dangereuse et doivent être protégés

Installation and Wiring

de façon adéquate de tout contact de l’utilisateur hors du boîtier du modèle

5200A.

Perform steps 1 - 4 of 5200A installation.

1.

Open front panel -

2.

page 39.

Feed relay device wires through drilled hole in rubber grommet and to

3.

locations C. The terminal strip is removable.

CAUTION:

Run high and low voltage cables through separate bulkhead and

conduit.

Terminate relay device wires according to relay number. Wire devices to

4.

NC or NO and C pins depending on device configuration.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.19

Complete other wiring to 5200A.

5.

Close front panel -

6.

page 40.

Test relays.

7.

Wiring a Contactor

The four onboard relays are rated at 240VAC@5AMPs (110VAC@10AMPs) switch-

ing capacity. Contactors should be used for devices drawing current loads higher

than relay specification. See figure 3.20 for contactor wiring.

Installation and Wiring

Advertising