SDI-12 Command and Response Protocol 3-15

3.15 Extended Read Power_Mode and Write Power_Mode
The 4-20mA output is updated whenever a measurement is made. For industrial applications
where the H-3342 is connected to a SCADA or PLC system and low-power is not of concern, the
H-3342 can be programmed to make continuous measurements. This command is used to change
the power mode. The H-3342 comes from the factory with the power mode set to the Sleep mode.

Once a new value is written, a copy is sent to the sensor data buffer for verification. This data
can be viewed by using a subsequent "D" command. To read or verify the value any other time,
use the "XRPMD" command.

Command

Response

Description

"aXRPMD!"

“a0011<cr><lf>"

Read Power_Mode

"aXWPMDn!"

“a0011<cr><lf>"

Write Power_Mode

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XRPMD

are upper case characters.

XWPMD

are upper case characters.

n

is the new setting (0 or 1)
0 = Sleep between measurements
1 = Make continuous measurements (always on)

This command takes 001 seconds to complete and places 1 value in the data buffer. Use the
“aD0" command to collect and view the current value.

Example of a H-3342 Extended "Read Power_Mode" command:

Command

Response

Time

Values

Description

"aXRPMD!"

"a0011<cr><lf>"

1 sec

1

Read Power_Mode

Command

Response

Description

"aD0!"

"a+0<cr><lf>"

Mode = Sleep

Example of a H-3342 Extended "Write Power_Mode" command:

Command

Response

Time

Values

Description

"aXWPMD1!"

"a0011<cr><lf>"

1 sec

1

Write Power_Mode

Command

Response

Description

"aD0!"

"a+1<cr><lf>"

Mode = Always On