H-3531 – Xylem H-3531 FlashLite User Manual

Page 48

Advertising
background image

5-10 SDI-12 Command and Response Protocol

H-3531

5.9 Change Sensor Address Command
The Change Sensor Address Command allows the sensor address to be changed. The address is
stored in non-volatile EEPROM within the sensor. The H-3531 FlashLite™ will not respond if
the command was invalid, the address was out of range, or the EEPROM programming operation
failed.

Command

Response

Description

"aAn!"

"n<cr><lf>"

Change sensor address

Where:

a

is the current (old) sensor address ("0-9", "A-Z", "a-z", "*", "?"). An ASCII
"*" may be used as a "wild card" address if the current address is unknown
and only one sensor is connected to the bus.

A

is an upper-case ASCII character.

n

is the new sensor address to be programmed ("0-9", "A-Z").

NOTE: To verify the new address use the "Identify Command."

Example of a "Change Sensor Address" command:

Command

Response

Description

"aA2!"

"2<cr><lf>"

Change sensor address to "2"