What to do if...

ConOx

42

6

What to do if...

6.1

Error symptoms of DO measurement

6.2

Error symptoms of conductivity measurement

Error symptom

Cause

Remedy

The sensor is in the air
and the display shows
0.0 mg/l or 0 % O

2

– No connection between

measuring instrument and sensor

– Cable defective
– No electrolyte in the membrane

cap

– Check connection between

measuring instrument and sensor

– Replace and refill the membrane

cap (see section 5.3)

The sensor cannot be
calibrated

– Contaminated membrane cap
– Electrolyte depleted

– Replace and refill the membrane

cap (see section 5.3)
Subsequently, wait for 30 to
50 min and recalibrate.

The sensor still cannot
be calibrated after
changing the electrolyte
and membrane cap

– Contaminated electrodes or

sensor toxification

– membrane cap not screwed on

tight enough

– Clean the electrodes (see

section 5.4)

– Screw membrane cap tighter

Incorrect temperature
display

– The temperature sensor of the

conductivity basic module was not
immersed deep enough in the
measuring solution

– Temperature sensor defective

– Observe the minimum immersion

depth

– Return the sensor

Mechanical damage to
the sensor

– Return the sensor

Error symptom

Cause

Remedy

No temperature or con-
ductivity display

– No connection between

measuring instrument and sensor

– Cable defective

– Check connection between

measuring instrument and sensor

Measurement delivers
implausible conductivity
values

– Incorrect cell constant adjusted at

the measuring instrument

– Measuring range exceeded
– Contamination in the area of the

electrodes

– Electrodes damaged

– Check / correct the cell constant
– Observe the application range
– Clean the sensor (see

section 5.2).

– Return the sensor

Incorrect temperature
display

– The temperature sensor was not

immersed deep enough in the
measuring solution

– Temperature sensor defective

– Observe the minimum immersion

depth

– Return the sensor