Copper – LaMotte COD Plus Colorimeter 1922-EX2 User Manual

Page 82

Advertising
background image

COPPER

DIETHYLDITHIOCARBAMATE METHOD • CODE 3646-SC

QUANTITY

CONTENTS

CODE

15 mL

*Copper 1

*6446-E

*WARNING: Reagents marked with * are considered hazardous substances. Material
Data Safety Sheets (MSDS) are supplied for these reagents. For your safety, read label
and accompanying MSDS before using.

The copper content of drinking water generally falls below 0.03 parts per
million, but copper levels as high as 1.0 part per million will give water a bitter
taste. Waters testing as high as 1.0 part per million copper have probably been
treated with a copper compound, like those used in the control of algae, or
have become contaminated from untreated industrial wastes. The addition of
copper sulfate to lakes causes an increase in the copper content of the
sediments. Acid waters and those high in free carbon dioxide may cause the
corrosion or “eating away” of copper, brass and bronze pipes and fittings. This
corrosion results in the addition of copper into the water supply.
APPLICATION:

Drinking, surface, and saline waters; domestic and industrial
wastes.

RANGE:

0.00 – 6.00 ppm Copper

METHOD:

Cupric ions form a yellow colored chelate with
diethyldithiocarbamate around pH 9-10 in proportion to the
concentration of copper in the sample.

SAMPLE
HANDLING &
PRESERVATION:

Copper has a tendency to be adsorbed to the surface of the
sample container. Samples should be analyzed as soon as
possible after collection. If storage is necessary, 0.5 mL of
20% hydrochloric acid per 100 mL of sample will prevent
“plating out.” However, a correction must be made to bring
the reaction into the optimum pH range.

INTERFERENCES: Bismuth, cobalt, mercurous, nickel and silver ions and

chlorine (6 ppm or greater) interfere and must be absent.

COD PLUS TEST PROCEDURES 6/02

COPPER - CUPRIZONE 1/2

Advertising
This manual is related to the following products: