3 - converter setup 1 page3, 0,00 g/l – Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual

Page 25

Advertising
background image

25

FW: 841013-005/843003-001

Turbidity and Suspended Solids Transmitter

EN 5.10 SuSix Manual 1406

4. Proceed to “Sensor Calibration” in the “Converter Setup” menu (see

below).
Note that the options "Sensor Calibration", "Cleaning" and "Solid
curve" will be marked with an "NA" (not applicable), if they cannot be
selected (for example if wrong units have been chosen).

Averaging

Units

mA output

Converter Setup

Quality improvement

No

Yes

Back Select

>

20 mA output

20 mA =

20.00 g/l

Back Select

OK

Units

NTU (Turbidity)

FTU (Turbidity)

FNU (Turbidity)

EBC (Turbidity)

% (Suspended solids)

mg/l (Suspended solids)

g/l (Suspended solids)

Kg/l (Suspended solids)

Back Select

OK

Fortsætter på side 5

Sensor Calibration

Averaging

Units

mA output

Cleaning

Solid curve

High Alarm

Low Alarm

Sensor error

Status

Factory settings

Service menu

Back Select

OK

4mA output

4 mA =

00.00 g/l

Back Select

OK

accordingly

Sensor Calibration

Yes

Use mA output

Yes

No

Back Select

OK

Averaging

0001 sec

(from 1 sek to 120 min)

Back Select

>

Yes

Adjust calibration

Select point to adjust

1 point Done

2 point Adjust

3 point

Back Select

OK

Calibration point 1

Type in lab. Value

10,2 g/l

Ok (/ Next)

Adjust

Back Select

Ok

Adjust

Calibration point 1

010,230 g/l

Back Select

>

Adjust

1

point

Adjust

Calibration point 1

0,00 g/l

Back Select

>

Adjust samples

Adjust

1. point

00010,2 g/l

Back Select

>

Adjust sample

1 Measured

10,2 g/l

2 Measured

8,11 g/l

3 Measured

5,87 g/l

Back Select

OK

Next

Select point

Adjust
2. point

00008,11 g/l

Back Select

>

Adjust
3. point

00005,87 g/l

Back Select

>

Sensor Calibration

Measurement samples

Adjust samples

Clear ”Measurement sample”

Clear ”Adjust samples”

Back Select

OK

1./2./3. point

Start sample?

Progress

[ 0% ]

Exit Start

Measurement samples

Select sample

1. point Measured

2. point Measured

3. point Measured

Back Select

OK

Clear ”Adjust samples”

Clear ”Adjustment sample”

No

Yes

Back Select

Ok

Start point

2. point

1. point

3. point

Next

Ok

Ok

Ok

Select point

Select point

1./2./3. point

Measureing

Progress

[||||||||||||||||| ] 45%

Cancel

Measure point 1/2/3

Done

Progress

[ 100% ]

Exit Next

Select point

Clear ”Measurement sample”

No

Yes

Back Select

Ok

Clear ”Measurement sample”

3 - Converter setup 1

Page3

5. Select “Measurement samples”, “1. point” and close with OK.

Advertising