Xylem Oxi 3210 Dissolved oxygen (D.O.) meter User Manual

Page 29

Advertising
background image

Oxi 3210

Operation

29

ba75795e02

03/2009

4.4.3

Settings for D.O. sensors
(menu or measurement and calibration settings)

Overview

The following settings are possible for D.O. sensors:

z Salinity correction

z Salinity (salinity equivalent)

z Calibration interval

z Automatic Stability control

Settings

The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the <F1>/[menu] or <ENTER> key.
After completing the settings, switch to the measured value display with
<M>.

Menu item

Possible
setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.

Sal correction

On
Off

Manual salt content
correction for concentration
measurements.

Salinity

0.0 ... 70.0

Salinity or salinity equivalent
for the salt content
correct
ion.
This function is only
available for concentration
measurements if the
manual salt content
correction is switched on.

Stability control

On /Off

Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)